Visit our Customer Experience Center for all of our latest demos Register Now
OH0TA Medical Image Sensor for Endoscopes and Catheters Improves Image Quality and Reduces Power Consumption
Color CMOS Analog 160 Kpixel (400 x 400) Image Sensor with PureCel®Plus‑S Technology
  • SKU
    OH0TA10-A04A

Medical Image Sensor for Endoscopes and Catheters Improves Image Quality and Reduces Power Consumption

OH0TA Product Q&A Series Part 1 OMNIVISION’s OH0TA is an OVMed® medical grade image sensor with a package size of just 0.55 mm x 0.55 mm, featuring a 1.0 micron pixel and a 1/31” optical format. The OH0TA takes medical visualization to the next level by quadrupling the RGB image resolution of its predecessor to 400 x 400, or 160 Kpixels, at 30 frames per second, while reducing the power consumption by 20% to 20 mW. This allows designers to add ultra-compact visualization to single-use and reusable endoscopes, as well as catheters and guidewires, with a small outer diameter of 1-2 mm.

OH0TA Product Q&A Series Part 2 The OH0TA’s increased resolution improves the visualization of tissues and other fine details inside the human body, allowing surgeons to see crisper images during procedures such as neuro, ophthalmic, ENT, cardiac, spinal, urology, gynecology and arthroscopy, as well as dental and veterinary diagnosis and surgery. Additionally, the sensor’s lower power consumption reduces “chip on tip” camera heat for greater patient comfort and longer procedure durations, while reducing noise.

To achieve this increase in resolution, along with a smaller pixel size and optical format, the OH0TA is built on OMNIVISION’s PureCel®Plus-S stacked die technology. This next-generation pixel technology also provides higher color fidelity and excellent low light sensitivity of 3600 mV/lux-sec, along with a high signal-to-noise ratio of 37.5 dB for greater image quality. Additionally, PureCel®Plus-S enables the OH0TA’s higher full well capacity (FWC), zero blooming and lower power consumption, resulting in better patient outcomes.

Other key features include a 15.5 degree chief ray angle, enabling the use of lenses with high fields of view and short focus distances. It also supports a 4-wire interface, as well as raw analog data output, both of which can transmit via cables as long as 4 meters with minimal signal noise. The OH0TA sensor also interfaces with OMNIVISION’s existing OV426 analog-to-digital-conversion bridge chip for backward compatibility and easy adoption. Additionally, it is autoclavable for reusable endoscope sterilization.

March 9, 2021: OMNIVISION Quadruples Resolution and Reduces Size of World’s Smallest Commercial Image Sensor; Breaks Own Guinness World Record

Optical size of 1/31″

AntLinx™ Analog output

Single 3.3V power supply

On-chip PLL

Serial peripheral interface (SPI)

Exposure and gain control

Pseudo-global shutter (LED mode)

PureCel®Plus-S pixel structure

Improved sensitivity, FWC, zero blooming, low noise, and low power consumption

Enhanced NIR sensitivity

Square aspect ratio

Minimum package size (total 4 pads)

Autoclavable

4 m drive distance

  • Spec
    OH0TA10-A04A
  • Analog / Digital
    Analog
  • Category
    1MP and Below
  • CFA (Chroma)
    Color
  • Frame Rate
    Full @ 30 fps
  • Image Area
    411.264 × 411.264 μm
  • Interface
    Analog, AntLinx™, SPI
  • Optical Format
    1/31"
  • Output Format
    Analog
  • Package
    CSP
  • Package Dimensions
    550 x 550 μm
  • Pixel Size
    1.008 µm
  • Power Requirement
    25 mW (with IO consumption)
  • Resolution
    400 x 400
  • Shutter Type
    Rolling Shutter
  • Technology
    AntLinx™, PureCel®Plus‑S
  • Use Case
    Brain / Eyes / Heart / Spine - Angioscope, Brain / Eyes / Heart / Spine - Spinoscope, Dental Mirror, Ear / Nose / Throat and Airway Management - Esophagoscope, Ear / Nose / Throat and Airway Management - Otoscope, Ear / Nose / Throat and Airway Management - Rhinoscope, Musculo-skeletal - Arthroscope, Stomach and Intestine - Proctoscope, Urology and Gynaecology - Choledochoscope, Urology and Gynaecology - Cystoscope, Urology and Gynaecology - Hysteroscope, Urology and Gynaecology - Nephroscope, Urology and Gynaecology - Utero-renoscope
  • Operating Temperature
    -20°C to +70°C